Skip to product information
1 of 3

VOLUMETRIC INFUSION PUMP

VOLUMETRIC INFUSION PUMP

Regular price Rs. 27,000.00
Regular price Sale price Rs. 27,000.00
Sale Sold out
💬 Get a Quote on WhatsApp
Quantity

Glamatic GPV1108 – Volumetric Infusion Pump

Product Description

The Glamatic GPV1108 Volumetric Infusion Pump is a reliable and precise infusion solution designed for controlled delivery of intravenous fluids and medications in hospitals and clinical environments. It provides accurate flow-rate control, extensive infusion volume settings and multiple safety features to support safe and consistent infusion therapy.

The pump offers an adjustable flow rate from 1 mL/h to 1,200 mL/h, with flow-rate accuracy of ±5% after correct calibration and mechanical precision of ±2%. Its intelligent monitoring system continuously checks infusion conditions and provides audible and visual alarms for events such as occlusion, end of infusion, door opening, air bubbles, incorrect settings, low battery and AC power interruption.

An integrated ultrasonic bubble detector helps identify air bubbles in the infusion line, while adjustable occlusion pressure levels allow the device to respond appropriately to different infusion conditions. The built-in rechargeable lithium battery provides approximately 4 hours of backup operation, helping maintain infusion during temporary power interruptions.

With a compact design, IPX4 protection and Type CF applied-part classification, the GPV1108 is designed for dependable operation across a variety of clinical applications.

Key Features

  • Volumetric Infusion Pump

  • Model: GPV1108

  • Brand: Glamatic

  • Designed for hospital and clinical applications

  • Adjustable flow rate from 1–1,200 mL/h

  • Flow-rate accuracy: ±5% after correct calibration

  • Mechanical precision: ±2%

  • Adjustable total infusion volume

  • KVO function

  • Three-level adjustable occlusion alarm

  • Ultrasonic air-bubble detection

  • Audible and visual alarm system

  • Human voice alarm prompts

  • End-of-infusion detection

  • Door-open alarm

  • Air-bubble alarm

  • Wrong-setting alarm

  • Low-battery alarm

  • AC power interruption alarm

  • Built-in rechargeable lithium battery

  • Approximately 4 hours battery backup

  • Compact and lightweight design

  • IPX4 protection

  • Type CF classification

Infusion Specifications

Parameter Specification
Flow Rate 1–1,200 mL/h
Flow Rate Accuracy Within ±5% after correct calibration
Mechanical Precision Within ±2%
Pulse Rate 100–1,000 mL/h, 100 mL/h step
Total Infusion Volume 1–9,999 mL
Infusion Volume Setting 0.1–9,999.9 mL
KVO Rate 1–5 mL/h, 1 mL/h step

Occlusion Pressure

The GPV1108 provides three adjustable occlusion alarm levels to accommodate different infusion requirements.

Alarm Level Pressure
High 800 ± 200 mmHg
Medium 500 ± 100 mmHg
Low 300 ± 100 mmHg

Approximate pressure equivalents:

  • High: 106.7 ± 26.7 kPa

  • Medium: 66.7 ± 13.3 kPa

  • Low: 40.7 ± 13.3 kPa

Air Bubble Detection

The integrated ultrasonic bubble detector monitors the infusion line for air bubbles.

Parameter Specification
Detection Method Ultrasonic Wave
Detection Sensitivity ≥25 µL

This provides an additional layer of safety by detecting air bubbles during infusion.

Alarm & Safety System

The pump features an intelligent audible and visual alarm system, including human voice alarm prompts for important operating conditions.

Alarms include:

  • End of infusion

  • Occlusion

  • Door open

  • Air bubbles in the infusion tube

  • Incorrect setting

  • Low battery

  • AC power interruption

  • Other system operating conditions

These alarms help healthcare professionals identify potential infusion issues promptly.

Power Supply & Battery

Parameter Specification
AC Power Supply 100–240 V AC
Frequency 50/60 Hz
Battery Internal Rechargeable Li Battery
Battery Capacity ≥1,600 mAh
Battery Backup Approximately 4 hours
Power Consumption 18 VA
Fuse F1AL/250V, 2 pcs

The internal rechargeable battery provides backup operation during temporary power interruptions.

Operating Conditions

The GPV1108 is designed for operation under the following environmental conditions:

Parameter Specification
Ambient Temperature +5°C to +40°C
Relative Humidity 20–90%
Atmospheric Pressure 86.0–106.0 kPa

Transport & Storage Conditions

Parameter Specification
Ambient Temperature -30°C to +55°C*
Relative Humidity ≤95%

*Based on the supplied manufacturer specification.

Equipment Classification

Parameter Specification
Equipment Classification Class II
Power Supply Internal Power Supply
Applied Part Type CF
IP Classification IPX4

Physical Specifications

Parameter Specification
Dimensions 140 × 157 × 220 mm
Net Weight Approximately 1.8 kg

Applications

The Glamatic GPV1108 is suitable for controlled infusion applications in:

  • Hospitals

  • Intensive Care Units (ICU)

  • General Wards

  • Operation Theatres

  • Emergency Departments

  • Paediatric Departments

  • Neonatal Care

  • Clinics

  • Nursing Homes

  • Critical Care Units

  • Medical and Surgical Departments

Product Highlights

Glamatic GPV1108 | Volumetric Infusion Pump | 1–1200 mL/h Flow Rate | ±5% Flow Accuracy | ±2% Mechanical Precision | 1–9999 mL Infusion Volume | KVO 1–5 mL/h | 3-Level Occlusion Alarm | Ultrasonic Bubble Detector | ≥25 µL Detection Sensitivity | Human Voice Alarm | End-of-Infusion Alarm | Door-Open Alarm | Low-Battery Alarm | 4-Hour Battery Backup | 100–240 V AC | IPX4 | Type CF | 1.8 kg Compact Design

View full details
Your cart
Product Product subtotal Quantity Price Product subtotal
VOLUMETRIC INFUSION PUMP
VOLUMETRIC INFUSION PUMP
VOLUMETRIC INFUSION PUMP
Rs. 27,000.00/ea
Rs. 0.00
Rs. 27,000.00/ea Rs. 0.00